High Mass Resolving Power, High Mass Accuracy, Benchtop Mass Spectrometry Instrument
II.1.2) Prif god CPV
38000000
II.1.3) Y math o gontract
Cyflenwadau
II.1.4) Disgrifiad byr
The instrument must provide 480,000 mass resolving power, < 1ppm mass accuracy, a dynamic range of > 5000:1, scan rates up to 40 Hz and rapid polarity switching < 1 second per cycle. It must be supplied with an electrospray ion source.
II.1.6) Gwybodaeth am lotiau
Mae’r contract hwn wedi’i rannu’n lotiau:
Na
II.1.7) Cyfanswm gwerth y caffaeliad
Gwerth heb gynnwys TAW: 800 000.00 GBP
II.2) Disgrifiad
II.2.3) Man cyflawni
Cod NUTS:
UK
II.2.4) Disgrifiad o’r caffaeliad
Our large-scale instrument development mass spectrometry imaging (MSI) studies require high mass accuracy (<1 ppm), high mass resolving power (> 400,000) and the ability to couple to a range of third party and home-built atmospheric pressure developmental ion sources. These home-built ion sources were developed for coupling to a Thermo Orbitrap Elite instrument. This package must also have a small footprint, preferably benchtop, profile to fit into our laboratory space. Currently, of commercially available MSI instruments, the Thermo Orbitrap Exploris 480 is the only instrument to fulfil these requirements completely.
II.2.11) Gwybodaeth am opsiynau
Opsiynau:
Na
II.2.13) Gwybodaeth am Gronfeydd yr Undeb Ewropeaidd
Mae'r broses gaffael yn gysylltiedig â phrosiect a/neu raglen a ariennir gan gronfeydd yr Undeb Ewropeaidd:
Na
Section IV: Gweithdrefn
IV.1) Disgrifiad
IV.1.1) Y math o weithdrefn
Dyfarnu contract heb gyhoeddi galwad am gystadleuaeth yng Nghyfnodolyn Swyddogol yr Undeb Ewropeaidd ymlaen llaw
Cyfiawnhau'r weithdrefn ddyfarnu a ddewiswyd:
Mae’r caffaeliad y tu allan i gwmpas cymhwysiad y gyfarwyddeb
Esboniad:
The Thermo Orbitrap Exploris 480 is the only commercially available mass spectrometry instrument providing the following required combination of characteristics:
• Scan rates up to 40 Hz
• Benchtop, small footprint instrument.
• Compatibility with third party MALDI MSI and LESA ion sources and our existing home-built atmospheric pressure ion sources.
• The mass spectrometer is a high-resolution FT system.
• Achievable mass resolution must be greater than 400,000 FWHM at m/z 200 for all scan modes and both polarities.
A copy of the Regulatory Compliance documentation (CE or UKCA Declaration of Conformity) certificate must be provided with the instrument, which will be reviewing by NPL’s regulatory compliance expert upon receipt.
IV.1.8) Gwybodaeth am Gytundeb Caffael y Llywodraeth (GPA)
The procurement is covered by the Government Procurement Agreement:
Na
Section V: Dyfarnu contract/consesiwn
Teitl: High Mass Resolving Power, High Mass Accuracy, Benchtop Mass Spectrometry Instrument
V.2 Dyfarnu contract/consesiwn
V.2.1) Y dyddiad y daeth y contract i ben/consesiwn
02/06/2023
V.2.2) Gwybodaeth am dendrau
Dyfarnwyd y contract i gr?p o weithredwyr economaidd:
Na
V.2.3) Enw a chyfeiriad y contractwr
Thermo Fisher
Stafford House, Boundary Way
Hemel Hempstead
HP2 7GE
UK
NUTS: UK
BBaCh yw’r contractwr:
Na
V.2.4) Gwybodaeth am werth y consesiwn a’r prif delerau cyllido (heb gynnwys TAW)
Cyfanswm gwerth y consesiwn/lot: 800 000.00 GBP
V.2.5) Gwybodaeth am is-gontractio
Mae’n debygol y /consesiwn caiff y contract ei is-gontractio